ESTREAM®
Automated PCR setup from easyMAG® eluates
The ESTREAM instrument is a compact pipetting instrument for any laboratory requiring fully automated setup of real-time PCR amplification reactions and sample traceability.
The system can prepare a 96-well PCR reaction plate from nucleic acid eluates in either easyMAG vessels or storage tubes, or a combination of both, in ~35 minutes.
ESTREAM Overview
The ESTREAM instrument is a compact pipetting instrument for any laboratory requiring fully automated setup of real-time PCR amplification reactions and sample traceability. The system can prepare a 96-well PCR reaction plate from master mixes and nucleic acid eluates presented in either easyMAG vessels or storage tubes, or from both, in about 35 minutes.ESTREAM is designed to prepare multiple quantitative and qualitative real-time PCR assays. Intelligent pipetting from a test request file allows processing of samples requiring varying subsets of master mixes without re-programming the instrument.
The system includes a barcode reader, UV light for decontamination, and provides a positive pressure HEPA-filtered atmosphere.
ESTREAM Uses
- ESTREAM is designed to prepare multiple quantitative and qualitative real-time PCR assays.
- Enables more productivity from your personnel
- Less demanding workflow for your staff
- Frees prep time that can get the thermocycler started earlier
- Provides traceability to support accreditation and peace of mind
Supported thermocyclers and associated consumables | |
Life Technologies (ABI) | 7500: MicroAmp® Optical 96-Well reaction Plate 7500: FastMicroAmp® Fast Optical 96-Well Reaction Plate |
Bio-Rad Laboratories | CFX96 Real-Time: Hard-Shell® Low-Profile Thin-Wall 96-Well Skirted PCR Plates |
Roche | LightCycler 480 II: LightCycler® 480 Multiwell Plate 96, white |
QIAGEN | Rotor-Gene Q: Rotor-Disc 72, Strip Tubes and Caps, 0.1 ml, PCR Tubes, 0.2 ml |
ESTREAM Features
- Combine multiple quantitative and/or qualitative real-time assays per plate
- Supports frequently-used thermocycler consumables from ABI, Roche, BioRad and Qiagen
- Software that thinks like you do: combine samples plus master mixes to create reactions
- Intelligent pipetting from .csv files: daily workflow involves no re-programming
- Flexibility to run commercial assays and your laboratory-developed tests
- Magnetic racks hold easyMAG eluates stable during pipetting
- UV sterilization: UV light system for desk sterilization under closed lid conditions
- HEPA-Filter: High Efficiency Particulate Air Filter system provides positive pressure
- Cooled Reagent and PCR Plate Holders
ESTREAM Specifications
- Compact benchtop automated liquid handler
- 23.6” W x 20.3” D x 18.5” H
- Disposable aerosol-barrier tips, 50 μl and 200 μl
- 5 - 190 μl pipetting range
- Single channel, air filled
- Uses magnetic racks that hold ESTREAM eluates stable during pipetting.
- Up to 6 easyMAG vessels = 48 input eluates/per plate (in standard configuration)
- 24 passively-cooled positions for master mixes or controls
- Cooled position for PCR reaction plate
- Additional 64 tube positions for additional input elutes or destination storage tubes
- Supports frequently-used thermocycler consumables from ABI, Roche, BioRad and Qiagen
Parameters | ESTREAM | |
Year Introduced to Market | 2014 (Europe); 2015 (US) | |
Technology Type | Air based pipetting system + pLLS (pressure Liquid Level Sensing) | |
Instrument Footprint (W x D x H) |
Width = 600 mm - Depth = 515 mm - Height = 470 mm | |
Weight | 45 Kg | |
Operating Ambient Humidity | Relative humidity : 20 to 80% | |
Operating Temperature Range | 18 to 25°C | |
Electrical | 100-250 VAC, 50/60 Hz, 60 watts | |
Maximum Altitude Installation | 1000 m | |
Noise Output in Decibels (dB) | 42,5 dB(A) | |
Table Capacity | 9x microplates full positions and 3 reagent racks positions | |
Pipetting head | Single channel , exchangeable head | |
Pipetting tips | 96 racked filtered tips : 50µl, 200 µl | |
Accuracy | Single dispense (50µl tips; dry_well transfer): | 5µl from 1.5 mL Microtubes (Trueness < 6% ; CV < 6 %) |
10μl from 1.5 mL Microtubes (Trueness < 5% ; CV < 3 %) | ||
10 μl from easyMAG vessels (Trueness < 1% ; CV < 3 %) | ||
Single dispense (200µl tips; dry_well transfer): | 180 μl from 1.5 mL Microtubes (Trueness < 2% ; CV < 1 %) | |
Multi dispenses (200µl tips; dry_well transfer): | 15 μl from ARGENE Master mix tubes (Trueness < 1% ; CV < 3 %) (specific pre-characterized Sarstedt tubes) | |
Master mix on board | Up to 3 x 8 cooled reagents/control tube positions. For ARGENE assays or similar, up to 3 quantitative assays, up to 6 qualitative assays | |
Time to Result (minutes) | 35 min (depending of the number of master mixes used) | |
Walkaway Time in Minutes | 30 min | |
On-board Sample Capacity | 96 eluates | |
Add Samples with no Pause | No | |
Minimum / Maximum Sample Volume | From 15 to 1500 µl | |
Type of Tubes accepted | easyMAG vessels, 0,2 to 1.5 mL microtubes | |
Dead Volume | easyMAG vessels: < 10µl | |
master mix Simport tubes: < 40µl for multidispense | ||
QS Simport tubes: <15 µl | ||
Pre-characterized & approved microtubes 1.5 ml: < 15 µl | ||
STAT Handling Method | No | |
Carryover/Contamination Prevention | Force Tips change option | |
HEPA Filter | ||
UV Lamp | ||
Sample/PCR mixes Clot Detection | Not applicable (eluate) | |
Sample Liquid Level Sensing | Yes | |
On-board Reagent Refrigeration | No – Cooling blocks available for certain reagent vessels and PCR microplates | |
Auto Calibration | No | |
Software Inventory Management | No | |
Automatic Maintenance Procedure | No | |
Daily Maintenance (min) | < 10 min | |
Technical Support via Modem or Intranet | Possible (Vilink) | |
Remote Diagnostics | No | |
On-Line User's Manual | Yes (bioMérieux technical library) |
ESTREAM Performance
ESTREAM vs. manual pipetting comparison
Quantification Standard EBV (5x104 copies/reaction)
Manual vs. ESTREAM pipetting comparison